5 Calibration - Using an External Calibration Weight:

Calibration is required when the weighing machine is initially installed or if the scale is moved to a
substantial distance from the original location. 10 minutes warm up of scale is always needed before
calibration.

Turn the scale on and then turn it off

Press and hold ZERO key, and then press ON/OFF key to turn on the scale, the display will show
CAPu=

Press PCS key, display will show CAP. Press PCS key again, display will show CALu=, press
UNIT key to choose the calibration unit kg or lb

Press PCS key to confirm and display will show CALu= again

Press PCS key, display will show CAL, press UNIT key to set the calibration weight, display will
show xx.x with a flashing digit

Press UNIT key to move flashing digit to right, press ZERO key to increase the digit number. (We
suggest the calibration weight to be at least 50% or more within the scale capacity to get an
accurate weighing )

Press PCS to confirm the setting, display will show CAL

Press ZERO key to start calibration, the display will show the AD value, and wait for stable
indicator displayed, then press UNIT key, display will show ------, and then the flashing calibration
weight

Place the test weight onto the center of the weighing pan, and wait for stable indicator displayed,
press UNIT key, after stable indicator displayed again, it will show ------ and then the AD value,
and now the calibration is completed

Remove the test weight and press ON/OFF key to turn off the scale. Turn it on again to see if the
weighing is accurate, if not, repeat above steps .