WEIGHING

Before weighing

.
Error messages

Err-O: Overload

: Low battery

Calibration may be required before weighing.

Read "CALIBRATION" first and if necessary, calibrate your scale/balance for

accurate weighing.

Weighing procedures

1.

Press [ON/OFF] to turn on the scale/balance.

When power is turned on, all display segments appear on LCD for a few
seconds and then finally "0" will appear on the display.

2.

Select the weighing unit with [UNIT].

Press [UNIT] to select a weighing unit, eg. "kg"、"g"

……

Once the unit has been selected, the unit arrow will be displayed next to the

weight value.

3.Start weighing

If you do not use a container for weighing

Verify the reading is "0". If not, press [ZERO] to display "0".

Place objects on the weighing platform to weigh.

When the reading becomes stable, the stable indicator is displayed.

If you use a container for weighing

Place an empty container on the platform.

Wait for the stability indicator to be displayed and press [ZERO].

Place the objects to be weighed in the container.

When the reading becomes stable, the stable indicator is displayed.

3

CALIBRATION

When to calibrate

Calibration may be required when it is initially installed, if the scale/balance is

moved to a substantial distance.

This is necessary because the weight of a mass in one location is not

necessarily the same in another location. Also, with time and use, mechanical

deviations may occur.

How to calibrate

Before entering calibration mode, There is a switch key on the bottom of the

scale. Turn on the switch key first, then press [ON/OFF] to turn the power on

for 1 minutes.

1. Press [ON/OFF] to turn the power off. Press [ZERO] first, then Press

[ON/OFF], while keep pressing [ZERO], the display will show CALU=, press

[UNIT] key to select full capacity unit kg, g,lb,oz. Press [ZERO] the display

will show XX.XXXX.

2. Press [UNIT] key to move the flash digit to right, press [PCS] key to increase

the flash digit. Ppress [ZERO] key to confirm the selection,.the display will

show "CAL" and then the A/D value ( the calibration weight should be more

than 2/3 of full capacity).

3. Press [UNIT], the display will show "-------" and then the calibration weight .

Place the calibration weight on platform. Press [UNIT] key , the display will

show "-------" and then the A/D value, Now calibration is finished. Remove the

weight from the platform.

4. Press [ON/OFF], turn off the power, press [ON/OFF] to turn on the power

again, place full weight on the platform to make sure whether weighing is

correct, if not, repeat steps 1-4.

5. Return to weighing mode

Press [ON/OFF] to turn the power off. Press [ON/OFF] again to turn on the

power and the scale/balance return to the weighing mode.