17 speechmap, Speechmap facts, Dsl 5.0 in speechmap – Audioscan Axiom User Manual

Page 76

Advertising
background image

Axiom

®

User's Guide Version 1.8

May 2015

5. Set Instrument to [CROS].

6. Select REAR 2 and obtain a response using [Pink noise] at 55. Adjust the response to match that obtained in

step 3.

7. Position the speaker directly in front of the client. Select REAR 3 and use [Pink noise] at 55 to check for a

smooth response. Repeat with Instrument set to [BTE]. An irregular response may be the result of reflections
from nearby objects (including the operator) or it may indicate phasing problems in the aids.