ROUTINE MAINTENANCE

The Activox DUO2 System was designed to minimize the amount of
routine maintenance that is required. The device includes no end-
user repairable parts. Aside from the preventative maintenance
outlined below, all other required maintenance must be performed
by qualified personnel.

INSTALLING A NEW AIR FILTER:

It is recommended that the Air Filter be changed yearly. However,
if the Activox DUO2 Stationary Base is used in dusty environments,
filter replacement may be required more frequently. The system
comes equipped with one factory-installed air filter and one
replacement filter. Contact your equipment provider for additional
replacement filters.

18

1. Flip the Power Switch to the “Off”

position.

2. Unplug the stationary base from the

electrical outlet.

3. Remove the Air Filter Access Door.

4. Remove the used Air Filter.

A. Wedge the pointed end of the

Maintenance Tool under the Air
Filter.

B. Raise the Air Filter up until it

dislodges from its position.

C. Pull the Air Filter out of its housing.