Maintenance


Fault finding

Fault symptom

Diagnosis

Action

Unit will not switch on

No/dead batteries fitted

Try another battery set (safely)

Unit will not switch off

Software locked out

Remove & replace battery clip
(safely)

Battery symbol appears

Battery state low

Replace batteries (safely)

Back light drains battery
life

Excessive use of backlight

Avoid or reduce use of backlight

Unit will not zero

Software locked out by
hardware

Remove & replace batteries (safely)

LCD appears blank/dim

Contrast control setting wrong

Turn off & on; press & hold contrast

Bar graph/number do not
change

Peak hold selected

Press zero or change Peak Hold
state

Range does not change

Range not set in Auto-range

Press & hold Range until Auto-range

He range will not show –5 Unit with nozzle ON and/or

range

Reset via Nozzle and Range keys

Gas group is wrong

Gas Group not set correctly

Press & hold Gases until shown OK

Nozzle/Peak not showing

Nozzle/Peak is not set
correctly

Press & hold Nozzle until both
correct

Calibration due message

Unit has exceeded 1 year on
cal

Return to Ion Science for
recalibration

Unit will not zero correctly Sensor contaminated

Leave to stabilise

Blocked probe

Probe blocked by solid or
liquid

Remove probe; clean; check now OK

Stalled motor

Motor failed & probe not
blocked

Return to supplier for repair

Instrument readings noisy Loose probe

Tighten nut slightly

Instrument readings noisy Test environment

contaminated

Move test into clean area

5000is will not switch on

Internal fuse blown

Return to supplier for repair

Table 5 – Fault finding