MINIPID 3PIN MANUAL

Ion Science Ltd

Page 17 of 27

Unrivalled Detection. www.ionscience.com

Intrinsic Safety



Schematic Block diagram

Intrinsic Safety circuit implementation

It is very important to abide by the stated temperature, power, voltage and current ratings.
This product is designed to drop into a standard LEL sensor position, however:

LEL sensors take considerable current and are often zoned by a separate 125 mA fuse and other suitable
upstream voltage limiting devices. Depending upon the current required by the monitoring electronic circuits
the MiniPID may either share the same zoned 125 mA fuse or the electronics can be located in another
zone whose power is supplied by another fuse.

If two zones are required, then very low current signals may be passed between the two zones by isolating
resistors to limit any potentially shared high current between the two zones thus maintaining separate zone
integrity.