Led’s, Selector pins, Calibration mode – Ion Science TVOC User Manual

Page 17: Calibration gases

Advertising
background image

TVOC MANUAL

Ion Science Ltd

Page 17 of 32

Unrivalled Detection. www.ionscience.com

Operation


LED’s

see Diagram 12
The green

indicates TVOC’s working status:

ON

Indicates the PID lamp is OFF

Flashing

Indicates the PID lamp is ON

A red indicates fault status:

ON

Indicates TVOC cannot function correctly. The 4-20mA output drops
to 3.5mA.
URGENT attention is required.

Flashing

Indicates the TVOC cannot function but a reading is being
attempted

A yellow Zero LED indicates during normal operation

ON

Last calibration zero level was outside expected limits

(The instrument will operate normally with the previous calibration)

A yellow Span LED indicates during normal operation

ON

Last calibration Span level was outside expected limits

(The instrument will operate normally with the previous calibration)


last calibration carried out was outside expected limits. The instrument will continue

working but service is required.

Please see Section 7 Calibration for YELLOW LED status.

Selector pins
TVOC has a range of options that are selected via the selector pins on the back of the main PCB (please see
the

‘TVOC Set up’ Section).

The options are shown below with the defaults highlighted in bold:

1-1000ppm

or

0.01

– 10 ppm or 0.1 – 100 ppm


ppm units

or

mg/m

3

units


Calibration mode
TVOC calibration mode can be accessed using the calibration
magnet supplied in the TVOC calibration tool kit (Part number: A-
845214).

To enter calibration place the magnet over the top right of the Ion
Science Logo until the state changes (i.e. an LED illuminates and/or
the display changes).

Ensure the magnet is withdrawn away from the magnetic switch by
at least 1 cm once a change has occurred.

Calibration gases
TVOC has three (3) ranges that can be selected however the 0-
10ppm range requires the instrument is calibrated by the user before
use. The 0-100ppm and the 0-1000 ppm ranges can initially run
using the Factory calibration carried out during the instruments
manufacture, however because the 0-10 ppm range is more
sensitive and more susceptible to environmental and system
tolerances the instrument must be calibrated before use.

If the 0-10 ppm range is selected TVOC will display the following
screen to indicate that a calibration is required before it can be used.

Follow the calibration instructions before use.

Diagram 10

Advertising