8.4.12
Status Commands
The status commands are used for querying and setting the state of the instru-
ment registers. For details on the operation of the status registers refer to the
IEEE-488.2 and SCPI standards.
8.4.12.1
*CLS
Clears the status registers.
*CLS
8.4.12.2
*ESE?
Returns the Event Status Enable Register.
*ESE?
255
8.4.12.3
*ESE <num>|MIN|MAX|DEF
Sets the Event Status Enable Register.
*ESE 48
8.4.12.4
*ESR?
Returns the Event Status Register.
*ESR?
0
8.4.12.5
*SRE?
Returns the Service Request Enable Register.
*SRE?
0
8.4.12.6
*SRE <num>|MIN|MAX|DEF
Sets the Service Request Enable Register.
*SRE 32
8.4.12.7
*STB?
Returns the Status Byte Register.
*STB?
200