Characteristic Values

1)

The device leakage current step of the test sequence is executed by means of

differential current measurement for testing per DIN VDE 0751.

2)

As of 25 mA: shutdown by residual current measurement within 100 ms

3)

Measured value P and calculated value S are compared and the smaller of the two is displayed.

4)

Measuring circuit becomes highly resistive, indication at display

5)

For DIN VDE 0701/0702: 2 k

Ω, for DIN VDE 0751: 1 kΩ

6)

This measuring range with DIN VDE 0751 only

Func-

tion

Measured Quantity

Measuring Range /

Nominal Range of

Use

Resolu-

tion

Nominal

Voltage

U

N

Open-Cir-

cuit Volt-

age U

0

Nominal

Current

I

N

Short-
Circuit
Cur. I

K

Internal

Resis-

tance R

I

Reference

Resist.

R

REF

Measuring

Error

Intrinsic Error

Overload Capacity

Value

Duration

Te

sts

p

er

DIN

V

DE

0701

/

0702

/

0751

Device Protective

Conductor

Resistance R

PE

0.000

… 2.100 Ω

1 m

Ω

4.5

… 9 V

DC

>200 mA

DC

±(5% rdg. + 10 d)

> 10 d

±(2.5% rdg. + 5d)

> 10 d

253 V

cont.

2.11

… 31.00 Ω

10 m

Ω

0.050

… 1.500 MΩ

1 k

Ω

50

… 500

V DC

1.0 • U

N

1.5 • U

N

> 1mA

< 10 mA

±(5% rdg. +10 d)

±(2.5% rdg. +5 d)

> 10 d

253 V

cont.

Insulation Resistance

R

ISO

1.01

… 10.00 MΩ

10 k

Ω

10.1

… 310.0 MΩ 100 kΩ

±(10% rdg. +10 d)

±(10% rdg. +10d)

Equivalent Leakage

Current

I

EL

or. I

DL

0.00

… 21.00 mA

10

μA

230 V~

– 20/

+10 %

< 3.5 mA > 72 k

Ω 1/2 kΩ

5

±(5% rdg. +10 d)

±(2.5 % rdg. +5 d)

> 10 d

253 V

cont.

20.1

… 120.0 mA

100

μA

Equivalent Patient
Leakage Current

I

EPL

0.0 ... 310.0

μA

100 nA

230 V~

– 20/

+10 %

< 3.5 mA > 72 k

Ω

1 k

Ω

±10 Ω

±(5% rdg. +10 d)

±(2.5% rdg. +5 d)

> 10 d

253 V

cont.

0.300 ... 2.100 mA

1

μA

2.00 ... 11.00 mA

10

μA

Contact Current or

Housing Leakage

Current I

Probe

or I

DL

0

… 310 μA

6

0.300

… 3.500 mA

0.1

μA

1

μA

1/2 k

Ω

5

±(5% rdg. +10 d)

±(2.5% rdg. +5 d)

> 10 d

253 V

cont.

Patient Leakage

Current I

PL

AC/DC Compo-

nents Measured

Separately

0.0 ... 310.0

μA

100 nA

1 k

Ω

±(5% rdg. +10 d)

±(2.5% rdg. +5 d)

> 10 d

253 V

cont.

2 4

0.300 ... 3.100 mA

1

μA

3.10 ... > 15.00 mA

10

μA

Differential

Current

ΔI

between L and N

1

0.000

… 3.100 mA

~

3.00

… 31.00 mA

~

2

1

μA

10

μA

±(10% rdg. +10 d)

> 10 d

±(5% rdg. +5 d)

> 10 d

2

2

Function T

e

st

Line Voltage U

L–N

207.0

... 253.0 V

~

0.1 V

±

(2.5% rdg. +5 d)

253 V

cont.

Load Current I

V

0 ... 16.00 A

RMS

10 mA

±

(2.5% rdg. +5 d)

20 A

10 min.

Active Power P

0 ... 3700 W

3

1 W

±

(5% rdg. +10 d)

> 20 d

253 V

cont.

20 A

10 min.

Apparent Power S

0 ... 4000 VA

1 VA

Calculated value U

L–N

• I

V

±

(5% rdg. +10 d)

> 20 s

Power Factor LF,

sinusoidal: cos

ϕ

0.00 ... 1.00

0.01

Calculated value P / S, display > 10 W

±

(10% rdg. +5 d)

Differential Current

Δ

I between L and N

per DIN VDE 0702

0.00 ... 31.00 mA~

10

μ

A

±

(10% rdg. +10 d)

> 10 d

±

(5% rdg. +5 d)

2

2

U

Probe

Probe voltage

0 ... 253.0 V

,

and

0.1 V

±

(2.5% rdg. +5 d)

> 10 d

253 V

cont.