Marathon OXYMITH TRANSMITTER F200060 User Manual

Page 22

Advertising
background image

Page 19 of 23

11/14/2006

Rev. 14

Thermocouple Null
Thermocouple Span
Cold Junction Trim

Communications port RS-485 Half Duplex Only

Protocol

Modbus RTU

Baud rates

1200, 2400, 4800, 9600, 19.2K (19.2K default)

Parity

None

Address

1 – 254 (Address 1 is default)

Housing

Material

Polyamide PA non-reinforced

Inflammability

Evaluation Class V0 (UL94)

Temperature Range

-40 to 100°C

Dielectric Strength

600 kV/cm (IEC243-1)

Mounting

Snaps on to EN 50022 top hat (T) style DIN rail.

Terminals

Wire clamp screw terminals on four position removable terminal blocks.
Wire Size

AWG 24 – 12 flexible stranded, removable terminal blocks.

Max. Torque

0.8 Nm

CAUTION: DO NOT CONNECT OR DISCONNECT HOUSING PLUGS
WHILE MODULE IS POWERED OR UNDER LOAD.

Weight

10 oz

Environmental Conditions

Operating Temperature

-20 °C to 55 °C (-4 to 130 F)

Storage Temperature

-40 °C to 85 °C (-40 to 185 F)

Operating and Storage Humidity

85% max relative humidity, noncondensing, from –20
to 65°C

Certifications and Compliance (PENDING)

Safety

EN 61010-1, IEC 1010-1

Safety requirement for electrical equipment for measurement, control, and
laboratory use, Part 1

Electromagnetic Compatibility
Immunity as specified by EN 50082-2

Electrostatic discharge

EN 61000-4-2

Level 3: 8 kV air

Electromagnetic RF fields

EN 61000-403

Level 3: 10 V/m
80 MHz – 1 GHz