INITIAL STARTUP OF THE CONCENTRATOR

NOTE: Concentrator may be used during the initial start warm-up time

(approximately 30 min.) while waiting for the O

2

purity to reach maximum.

When the unit is turned on, the green light will come on (SYSTEM OK / O

2

greater than

85%). After 5 minutes, the oxygen sensor will be operating

normally and will control the indicator lights depending on oxygen concentra-

tion values. The explanation of the indicator light functions are as follows.

OPERATING INSTRUCTIONS

SENSO

2

OXYGEN PURITY INDICATOR

(IF YOUR UNIT FEATURES THE O

2

SENSOR FEATURE)

This feature monitors the purity of oxygen generated by the oxygen

concentrator. If purity falls below factory preset standards, indicator

lights on the control panel will illuminate.

NOTE: The use of some accessories such as the pediatric flowstand and the

Homefill II compressor will deactivate the Low Flow Alarm.

WARNING

DO NOT change the L/min. setting on the flowmeter unless

a change has been prescribed by your physician or therapist.

NOTE: To properly read the flowmeter, locate the prescribed flowrate line on

the flowmeter. Next, turn the flow knob until the ball rises to the line. Now,

center the ball on the L/min. line prescribed (FIGURE 7).
2. If the flowrate on the flowmeter ever falls below 0.5L/min. for

more than about one minute, the LOW FLOW ALARM will be

triggered. This is a rapid beeping of the audible alarm. Check your

tubing or accessories for blocked or kinked tubing or a defective

humidifier bottle. After rated flow is restored, more than 0.5 L/

min., the LOW FLOW audible alarm will go off.

FIGURE 7 - ADJUSTING THE FLOWRATE-

PLATINUM 5 WITH SENSO

2

SHOWN

Ball

Flow Knob