Lifescan Pocket Scan Blood Glucose Monitor System User Manual
Page 84
MESSAGE
PROBLEM
WHAT TO DO
3. The meter was used in very
3. Move the meter away from
bright light.
the light source. Repeat the
test with a new test strip.
remove
The test strip was inserted
Repeat the test with a new test
strip
before the word
i n s e r t
strip. Wait for the word
insert
appeared on the display.
before inserting a test strip.
100 mg/dl
If the word
contrl
appears
Repeat the test with a new test
(Example)
after a blood test, your
strip and a larger drop of
contrl
blood sample was too small,
blood.
7 : 52am 7-01-00
smeared, or another drop was
added after the test began.
error 1
The sample was applied
Repeat the test with a new
retest
before the words
apply
test strip.
sample
appeared.
error2
1. The test strip moved
1. Repeat the test with a new
retest
during the test.
test strip.
77
MESSA