Installation 

MI 

33 

Go to the menu <Setup> <Sensor Setup>, and then select the register 
<Advanced Setup>. Use <Polling Address> for selecting the requested 
address. 

 

Figure 23: Address Setting 

Step‐by‐step instructions for addressing RS‐485 MI units: 

1.  Power the unit. 
2.  Using  either  the  buttons  or  software,  assign  unique  address 

to the sensor. 

3.  Power down the unit. 
4.  Repeat until all sensors have a unique address. 
5.  On  the  last  unit  in  the  network,  activate  the  shunt  resistor 

after the unit has been powered down. 

 

Failure  to  use  shielded  RS‐485  wire  or  activation  of  the 
shunt  resistor  when  the  unit  is  powered,  can  result  in 
damage to the electronics! 

 