Understanding out-of-range control test results – Акку-Чек Перформа Нано User Manual
Page 59
Understanding Out-of-Range Control Test Results
If the control result is not inside the acceptable range, here are some things you can
do to solve the problem:
Troubleshooting Checks
1. Check if test strips or control solutions
were expired.
2. Check if you wiped the tip of the control
solution bottle before and after use.
3. Check if the caps on the test strip
container and the control solution bottle
were always closed tightly.
4. Check if the test strip was out of the
container for a long time.
Action
Discard the test strips or control solutions
if either is past the expiry date. If the
control solution was opened more than
three months ago, discard it.
Wipe the tip of the bottle with a tissue.
Repeat the control test with a new test
strip and a fresh drop of control solution.
Replace the test strips or control solutions
if you think either may have been
uncapped for some time.
Repeat the control test with a new test
strip.
57
44702_EN_PNano_24_Ins.indd 57
44702_EN_PNano_24_Ins.indd 57
10/9/09 2:39:01 PM
10/9/09 2:39:01 PM