Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

EverGo is intended for use in the electromagnetic environment specified below� The user of EverGo should assure that it is used

in such an environment�

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF Emissions

CISPR 11

Group 1

The Device uses RF energy only for its internal

function� Therefore, its RF emissions are very

low and are not likely to cause any interference

in nearby electronic equipment�

RF Emissions

CISPR 11

Class B

The Device is suitable for use in all

establishments, including domestic

establishments and those directly connected to

the public low-voltage power supply network

that supplies buildings used for domestic

purposes�

Harmonic Emissions

IEC 61000-3-2

Class A

Voltage Fluctuations/Flicker Emissions

IEC 61000-3-3

Complies

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

EverGo is intended for use in the electromagnetic environment specified below� The user of EverGo should assure that it is used

in such an environment�

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic

Discharge (ESD)

IEC 61000-4-2

±6 kV Contact

±8 kV Air

±6 kV Contact

±8 kV Air

Floors should be wood, concrete or ceramic tile�

If floors are covered with synthetic material, the

relative humidity should be at least 30%�

Electrical Fast

Transient/Burst

IEC 61000-4-4

±2 kV for Power Supply

Lines

±1 kV for Input/Output

Lines

±2 kV for Power Supply

Lines

±1 kV for Input/Output

Lines

Mains power quality should be that of a typical

home or hospital environment�

Surge

IEC 61000-4-5

±1 kV Line to Line

±2 kV Line to Ground

±1 kV Line to Line

±2 kV Line to Ground

Mains power quality should be that of a typical

home or hospital environment�

Voltage Dips, Short

Interruptions and

Voltage Variations on

Power Supply Input

Lines

IEC 61000-4-11

<5% U

T

(>95% Dip in U

T

)

for 0�5 Cycle

40% U

T

(60% Dip in U

T

)

for 5 Cycles

70% U

T

(30% Dip in U

T

)

for 25 Cycles

<5% U

T

(>95% Dip in U

T

)

for 5 Seconds

<5 % U

T

(>95% Dip in U

T

)

for 0�5 Cycle

40 % U

T

(60% Dip in U

T

) for

5 Cycles

70 % U

T

(30% Dip in U

T

) for

25 Cycles

<5 % U

T

(>95% Dip in U

T

)

for 5 Seconds

Mains power quality should be that of a typical

home or hospital environment� If the user of the

Device required continued operation during

power mains interruptions, it is recommended

that the Device be powered from an

uninterruptible power supply or battery�

Power Frequency

(50/60 Hz) Magnetic

Field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a

typical home or hospital environment�

Note: U

T

is the A�C� mains voltage prior to application of the test level�