Ix71/ix51 – Olympus IX71 User Manual

Page 45

Advertising
background image

41

IX71/IX51

4

Attaching the Polarizer (IX-LWPO)

(Fig. 63)

# The polarizer can be mounted only on the IX2-LWUCD or IX2-

LWUCDA2.

}This polarizer has been designed for being mounted on a universal

condenser in case of DIC observation or simplified polarizer light
observation.

1. Loosen the polarizer clamping screw @ using the Allen screwdriver.
2. Align the pin hole on the condenser with the positioning pin ² on the

condenser, and insert the polarizer unit into the upper part of the condenser.

3. Tighten the polarizer clamping screw firmly.

1. Rotate the condenser’s turret for the BF (brightfield) light path (with no

optical element engaged in the light path). (Fig. 64)

2. Slide the polarizer detaching lever @ on the IX-LWPO polarizer to engage

the polarizer in the light path. (Fig. 64)

3. Engage the 10X objective in the light path, place an optimum specimen for

brightfield observation on the stage, bring the specimen into approximate
and remove the specimen out of the light path.

4. When the U-BI90CT binocular observation tube is used, rotate the CT

turret ² to position “CT” to engage the CT lens in the light path. (Fig. 65)
When the U-BI90/U-TBI90 binocular observation tube or U-TR30H-2
trinocular observation tube is used, remove an eyepiece and attach
the U-CT30 centering telescope.

5. Rotate the focus ring ³ (or, with the centering telescope, rotate the knurled

section) to bring the objective’s exit pupil into focus. (Fig. 65)

6. Slide the prism movement knob † of the DIC slider in the clockwise

direction around the axis until, the knob is stopped. A black interference
stripe then a rainbow-colored interference stripe will be observed. Here,
stop the knob at the position with which the black interference stripe can
be seen. (Figs. 61 & 66)

7. While observing the objective’s exit pupil, rotate the polarizer rotation/clamping

knob | on the polarizer unit horizontally until the black interference stripe
becomes darkest. This is the position for use of the polarizer. (Figs. 64 & 66)

8. After determining the position, clamp the polarizer rotation/clamping knob

so that the polarizer will not rotate.

Fig. 63

@

²

Fig. 64

|

Fig. 65

Fig. 66

5

Cross-Nicol Adjustment

(Figs. 64 to 66)

@

²

³

Advertising
This manual is related to the following products: