4 Mounting
VEGASOURCE 31 •
38131-EN-130415
Mounting brackets, angled attachments as well as mounting clips are
available for mounting.
Vertical pipeline, 30° inclined, diameter 50 … 100 mm
(1.97 … 3.94 in)
For pipeline diameters 50 … 100 mm (1.97 … 3.94 in), a diagonal
radiation path is recommended. The distance of the beam through the
medium is thus longer and an improved measuring effect is achieved.
For this, the optional lead shielding for the detector is recommended
in order to avoid influence from secondary radiation sources.
Fig. 13: 30° measurement arrangement on a pipeline with diameter
50 … 100 mm (1.97 … 3.94 in)
Vertical pipeline, diameter 50 … 420 mm (1.97 … 16.54 in)
For pipeline diameters 50 … 420 mm (1.97 … 16.54 in), a straight ra-
diation path is possible. The radiation-based sensor can be mounted
either horizontally or vertically.