CALIBRATION
CHAPTER 5
5-55
Using a laboratory millivolt meter, measure the voltage across pins 5 and 6.
Example: 08.0622 mV
7.
Program the measured voltage into the SURETEST Indicator as demonstrated in the example,
“+08.0622” by pushing and holding the Up or Down Cursor keys.
8.
Press ENTER. The display momentarily reads:
SAVING...
9.
Turn the power to OFF. Remove the calibration fixture.