Specifications

Operating Specifications

1

Range

0.20 ppb to 10 ppm

Precision

2

2% RSD

Accuracy

2

± 3%

Sample Type

On-line continuous; grab sample via the IOS System* (non-Turbo mode only)

Display Readout

3 significant digits

Calibration

Typically stable for 12 months

Analysis Time

4 seconds

t

5

2

2 minutes, 30 seconds

t

90

2

3 minutes, 30 seconds

Sample Temperature

1–95° C (34–203° F) – withstands short-term steam exposure

Ambient Temperature

10–40° C (50–104° F)

Sample Pressure

Up to 250 psi

Required Sample Line Flow Rate 30–300 mL/min (for on-line mode)

Instrument Sample Flow Rate 1.1 mL/min; 0.5 mL/min when TOC >5 ppm

Analyzer Specifications

900 Portable Turbo

900 On-Line Turbo

Outputs

Serial (RS-232) port (1); USB port (1);

4-20 mA output (1); binary output (1); serial

parallel printer port (1)

(RS-232) port (1); USB port (1); parallel printer

port (1); alarm outputs (2)

Display

Quarter-VGA, color touch-screen display Quarter-VGA, color touch-screen display

Power

Universal Power Supply: 100–240 ±10%

Universal Power Supply: 100–240 ±10%

VAC, 50 W, 50/60 Hz

VAC, 50 W, 50/60 Hz

Dimensions

H: 35.6 cm (14.0 in); W: 22.3 cm (8.8 in);

H: 62.4 cm (24.6 in); W: 45.2 cm (17.8 in);

D: 46.5 cm (18.3 in)

D: 26.4 cm (10.4 in)

Weight

12.5 kg (27.5 lb)

16.9 kg (37.2 lb)

Safety Certifications UL/cUL, CE

UL/cUL, CE

Industrial Ratings

IP-45 Enclosure

Consumables

UV Lamp

6 months

Acid Reagent

As needed, typically for 6 months (285 mL)

Oxidizer Reagent

As needed, typically 3-month stability; available in 150- or 300-mL cartridge

©2005, General Electric Company. All rights reserved.

300 00069 Rev E

MC05-128

* Trademark of General Electric Company; may be registered in one or more countries.
1 Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors

2 When TOC is less than 5 ppm

The Sievers 900 Series TOC Analyzers are protected by one or more of the following US and foreign patents: US 6271043; US 6228325; US 5976468; US 5902751;

US 5837203; US 5820823; US 5798271; US 5750073; US 5443991; US 5132094; EP 0897530; FR 0897530; GB 0897530; DE 69702516.0-08; EP 0471067. Other

patents pending.

The Americas

GE Analytical Instruments

6060 Spine Road

Boulder, CO 80301-3687 USA

T +1 800 255 6964

T +1 303 444 2009

F +1 303 527 1797

[email protected]

Europe/Middle East/Africa

GE Analytical Instruments

Unit 3, Mercury Way

Urmston, Manchester

UK M41 7LY

T +44 (0) 161 864 6800

F +44 (0) 161 864 6829

[email protected]

www.geinstruments.com

Asia Pacific

GE Analytical Instruments

42/F United Plaza

1468 Nanjing Rd (W)

Shanghai 200040

China

[email protected]

Advertising