3B Scientific Monocular Course Microscope Model 300 (230 V, 50__60 Hz) User Manual
Page 4
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany • www.3bscientific.com
Subject to technical amendments
© Copyright 2007 3B Scientific GmbH
Objectives: Revolver with 4 DIN achromatic objec-
tives 4x / 0.10, 10x / 0.25, 40x / 0.65, 100x / 1.25 oil
(with specimen protection)
Magnification: 40x, 100x, 400x, 1000x
Object stage: x-y cross table, 125 x 130 mm
2
, with
object guide and coaxial adjustment knobs per-
pendicular to the object stage, adjustment range
70 x 30 mm
2
Illumination: 5 W incandescent lamp incorporated
in the base integrated in base, with a converging
lens in lamp shaft, power supply 115 V resp. 230 V
50/60 Hz
Condenser: Abbe condenser N.A.1.25 NA 0.65 with
iris diaphragm , filter holder and blue filter, focus-
sed via rack and pinion drive
Dimensions: 220 x 154 x 359 mm³ approx.
Weight: 4.5 kg approx.
3. Unpacking and assembly
The microscope is packed in a molded styrofoam
container.
•
Take the container out of the carton remove
the tape and carefully lift the top half off the
container. Be careful not to let the optical i-
tems (objectives and eyepieces) drop down.
•
To avoid condensation on the optical compo-
nents, leave the microscope in the original pa-
cking to allow it to adjust to room tempera-
ture.
•
Using both hands (one around the pillar and
one around the base), lift the microscope from
the container and put it on a stable desk.
•
The objectives will be found within individual
protective vials. Install the objectives into the
microscope nosepiece from the lowest magni-
fication to the highest, in a clockwise direction
from the rear.
•
Insert the eyepiece into the tube.
4. Operation
4.1 General information
•
Set the microscope on a level table.
•
Place the object to be observed in the centre of
the specimen stage and clamp it to the object
guide.
•
Connect the mains cable to the net and turn on
the switch to get the object illuminated.
•
Make certain that the specimen is centered
over the opening in the stage.
•
Adjust the aperture of the iris diaphragm to get
the background brightness suitable for a high
contrast image.
•
Rotate the nosepiece until the objective with
the lowest magnification is pointed at the
specimen. There is a definite “click” when each
objective is lined up properly.
NOTE: It is best to begin with the lowest power
objective. This is important to reveal general struc-
tural details with the largest field of view first.
Than you may increase the magnification as nee-
ded to reveal small details.
When 100x (oil) objec-
tive is chosen, objective oil must be dripped onto
the slide.
To determine the magnification at which you are
viewing a specimen, multiply the power of the
eyepiece by the power of the objective.
•
Adjust the coarse-focusing-knob which moves
the stage up until the specimen is focused. Be
careful that the objective does not make con-
tact with the slide at any time. This may cause
damage to the objective and/or crack your
slide.
•
Adjust the fine-focusing-knob to get the image
more sharp and more clear.
•
Colour filters may be inserted into the filter
holder for definition of specimen parts. Swing
the filter holder out and insert colour filters.
•
Use the knobs of the mechanical stage to move
the slide side-, back- and forwards. The vernier
provides accurate location of the specimen
area.
4.2 Changing the fuse
•
Turn off the power switch and unplug the
mains plug.
•
Unscrew the fuse holder on the back of the
stand base with a screwdriver.
•
Replace the fuse and reinsert the holder in its
socket.
5. Storage and cleaning
•
Keep the microscope in a clean, dry and dust
free place.
•
When not in use always cover the microscope
with the dust cover.
•
Do not expose it to temperatures below 0°C
and above 40°C and a max. relative humidity
of over 85%.
•
Always turn off the light immediately after use.
•
Use a soft linen cloth and some ethanol to
clean the microscope.
•
Use a soft lens tissue to clean the optics.