2 measuring ranges, resolution, accuracy, Measuring ranges, resolution, accuracy – Xylem Lab 960 User Manual
Page 65
Lab 960
Technical data
65
ba75557e05
11/2009
Stand October 30, 2009
Translation of the legally binding German version
7.2
Measuring ranges, resolution, accuracy
Measuring ranges,
resolution
Variable
Measuring range
Resolution
ӕ
[µS/cm]
0.000 ... 1.999*
0.00 ... 19.99**
0.0 ... 199.9
0 ... 1999
0.001
0.01
0.1
1
ӕ
[mS/cm]
0.00 ... 19.99
0.0 ... 199.9
0 ... 500
0.01
0.1
1
Resistivity
[MOhm*cm]
0.00 ... 19.99
0.0 ... 199.9
0 ... 1999
0.01
0.1
1
SAL
0.0 ... 70.0
according to the IOT
table
0.1
TDS [mg/l]
0 ... 1999
Factor can be set
between 0.40 and
1.00
1
T [°C]
− 5.0 ... + 120.0
0.1
T [°F]
+ 23.0 ... + 248.0
0.1
* only possible with cells of the cell constant, 0.010 cm
-1
** only possible with cells of the cell constant, 0.010 cm
-1
or 0.090 ... 0.110 cm
-1
Cell constants
Cell constant C
Values
Can be calibrated in
the ranges
0.450 ... 0.500 cm
-1
0.585 ... 0.715 cm
-1
0.800 ... 1.200 cm
-1
Adjustable
0.250 ... 2.500 cm
-1
0.090 ... 0.110 cm
-1
0.010 cm
-1
(fixed)
Reference temperature
Reference
temperature
Values
Adjustable
20 °C (Tr20)
25 °C (Tr25)