Operation

handylab multi 12

42

ba75371e03

11/2009

4.5

Calibration intervals (Int 3, Int 4, Int 5)

For each measured variable, a time interval is stored. When it has expired,
you will be reminded to calibrate. After a calibration interval has expired, the
probe symbol of the relevant measured variable flashes. It is still possible to
measure. By calibrating the relevant probe, the function is reset and the in-
terval starts anew.

The following calibration intervals are set in the factory:

Setting the

calibration interval

You can change each of these intervals (1 ... 999 days):

Measured parameter

Designation

Default setting

pH/ORP

Int 3

7 days

Dissolved oxygen

Int 4

14 days

Conductivity

Int 5

180 days

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.

4

Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O

2

or

ӕ

) ap-

pears on the display. Example:

5

Set the required interval until the next calibration with <

>

<

>.

6

Confirm with <RUN/ENTER>.

7

Switch to the measuring mode with <M>.

°

C

1

TP

ARng

mg/l

LoBat

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

S

RCL

AR

Store

Auto

°

F

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

4

1

t 4

n

O

1

d

Time

Cal

Advertising