Physio-Control LIFEPAK 20e User Manual
Page 160

Specifications and Performance Characteristics
A-8
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
PACER
Pacing Mode
Demand or non-demand
Rate and current defaults (user configurable)
Pacing Rate
40 to 170 ppm
Rate Accuracy
±
1.5% over entire range
Output Waveform
Monophasic, amplitude stable to ±5% relative to leading edge
for currents greater than or equal to 40 mA, Duration 20
±1 msec, Rise/Fall times <= 1 msec [10-90% levels]
Output Current
0 to 200 mA
Pause
Pacing pulse frequency reduced by a factor of 4 when
activated
Refractory Period
200 to 300 msec
±
3% (function of rate)
ENVIRONMENTAL
Temperature, Operating
5° to 40°C (41° to 104°F)
Temperature, Non-
Operating
-20° to 60°C (-4° to 140°F) except therapy electrodes
Relative Humidity,
Operating
5 to 95%, non-condensing
Atmospheric Pressure,
Operating
Ambient to 522 mmHg (0 to 3048 meters) (0 to 10,000 ft)
Water Resistance,
Operating (without
accessories except for
ECG Cable and Hard
Paddles)
IPX1 (spillage) per IEC 60601-1 clause 44.6
EMC
IEC 60601-1-2:2001/EN 60601-1-2:2001, Medical Equipment-
General Requirements for Safety-Collateral Standard:
Electromagnetic Compatibility-Requirements and Tests.
IEC 60601-2-4:2002; Clause 36/EN 60601-2-4:2003: Clause
36, Particular Requirements for the Safety of Cardiac
Defibrillators and Cardiac Defibrillator monitors.
Shock (drop)
1 drop on each side from 45.7 cm (18 in.) onto a steel surface
Vibration
MIL-STD-810E Method 514.4, Cat 1