B&K Precision 1785B - Manual User Manual

Page 29

Advertising
background image

29


17. Reading product’s model, series number and version information (0x31)

1

st

byte

Start bit (0xAA)

2

nd

byte

Address (0x00~0xFE)

3

rd

byte

Command (0x31)

4

th

to 8

th

byte

Product model(ASIC code)

9

th

byte

Lower byte of the software version

10

th

byte

Higher byte of the software version

11

th

to 20

th

byte

Serial number(ASCII code)

21

st

to 25

th

byte

System reserve

26

th

byte

Check sum























18. Restore the factory default calibration data
(0x32)

1

st

byte

Start bit (0xAA)

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command(0x32)

4

th

to 25

th

byte

System reserve

26

th

byte

Check sum

19Enable the local key (0x37)

1

st

byte

Start bit ( 0xAA)

2

nd

byte

Address (0x00-0xFE)

3

rd

byte

Command (0x37)

4

th

byte

Enable/disable local key (0 is disable, 1is enable)

5

th

to 25

th

byte

System reserve

26

th

byte

Check sum code

Note:

For example, the serial number is 0123456789the product model is 6811and software

version is V2.03, then the returned data is as follows:

AA

00

31

36

38

31

31

00

03

02

30

31

32

33

34

35

36

37

38

39

XX

XX

XX

XX

XX

57



Here’s a conversation table between ASCII code and hexadecimal: