Introduction 3000ZA-3X


Teledyne

Analytical

Instruments

2

Three Zirconium oxide sensors to provide higher range

of analysis flexibility. The analyzer can cover ranges as
low as 0-10 ppm or as high as 0-95%.

No reference gas required

Versatile analysis over a wide range of applications.

Microprocessor based electronics: 8-bit CMOS

microprocessor with 32 kB RAM and 128 kB ROM.

Three user definable output ranges (from 0-1 ppm

through 0-95%) allow best match to users process and
equipment.

Auto ranging allows analyzer to automatically select the

proper preset range for a given measurement. Manual
override allows the user to lock onto a specific range of
interest.

Two adjustable concentration alarms and a system

failure alarm.

Extensive self-diagnostic testing, at startup and on

demand, with continuous power-supply monitoring.

CE compliance

RS-232 serial digital port for use with a computer or

other digital communication device.

Four analog outputs: two for measurement (0–1 VDC

and isolated 4–20 mA DC) and two for range
identification.

Convenient and versatile, steel, flush-panel or rack-

mountable case with slide-out electronics drawer.

1.4 Model Designations

3000ZA:

Standard model for sample under pressure

3000ZA-C: In addition to all the standard features, this

model also has separate ports for zero and
span gases, and built-in control valves. The
internal valves are entirely under the control
of the 3000ZA electronics to automatically