Paramagnetic O2 Purity Analyzer

Operation

Teledyne Analytical Instruments

43

• Setup

Auto-Cal

• Assign

Password

• Initiate

Self-Test

• Check

software

version

• View

sensor

output

• Log

out.

Zero

. Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine whether

each alarm will be active or defeated, HI or LO acting,
latching, and/or failsafe.

Range.

Used to set up three analysis ranges that can be

switched automatically with auto-ranging or used as individual
fixed ranges.

Any function can be selected at any time by pressing the appropriate

button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.

Each of these functions is described in greater detail in the following

procedures. The VFD screen text that accompanies each operation is
reproduced, at the appropriate point in the procedure, in a

Monospaced

type style. Pushbutton names are printed in

Oblique

type.

4.3 The System Function

The subfunctions of the

System

function are described below.

Specific procedures for their use follow the descriptions:

Auto-Cal:

Used to define an automatic calibration sequence

and/or start an Auto-Cal.

PSWD:

Security can be established by choosing a 5 digit

password (

PSWD) from the standard ASCII character set. (See

Installing or Changing a Password, below, for a table of ASCII
characters available.) Once a unique password is assigned and
activated, the operator MUST enter the UNIQUE password to
gain access to set-up functions which alter the instrument’s
operation, such as setting the instrument span or zero setting,
adjusting the alarm setpoints, or defining analysis ranges.