Passive Probe Compensation
Passive probes must be compensated to flatten overshoot. This is
accomplished by means of a trimmer at the connector end of the probe.
Attach the connector end of your passive probe to any channel.
Connect the probe end to the CAL output connector at the front of the
oscilloscope. Ground the probe.
Adjust the trim pot at the connector end of the probe until the square
wave is as flat as possible.
Display Dashboard
The WaveSurfer MXs-B oscilloscope’s display contains valuable information
about the current settings of your Vertical (channel), Horizontal
(Timebase), and Trigger controls. In addition, there are many shortcuts that
are available by using the touch screen capability of your display to quickly
access information or to open dialogs.