Actions Dialog
Various configurations can be made to all or some of your Pass/Fail
conditions using the Actions dialog.
By touching the Stop Test checkbox in the Actions dialog, you can set up the
test to end after a predetermined number of sweeps that you decide.
You can also decide the actions to occur upon your waveforms' passing or
failing, by selecting one or all of the following:
Stop
Audible alarm
Print image of display
Emit pulse (if available)
Save waveform
The Pulse checkbox causes a pulse to be output through the
Aux Out connector at the front of the oscilloscope. This pulse
can be used to trigger another oscilloscope. You can set the
amplitude and width of the pulse as described in Auxiliary Output Signals.
Depending on your oscilloscope model, you can configure up to 8 pass/fail
conditions. The Boolean conditions to determine if your waveform passes
are as follows:
Pass/Fail Boolean Conditions
All True
All False
Any True
Any False
All Q1 to Q4 Or All Q5 to Q8
Any Q1 to Q4 And Any Q5 to Q8