2 preparing the autopulse for use, 1 lifeband load-distributing band (ldb), 1 installing the lifeband – ZOLL AutoPulse Rev D Users Guide User Manual
Page 25: 2 preparing the autopulse for use -1, 1 lifeband load-distributing band (ldb) -1, 1 installing the lifeband -1, 2preparing the autopulse for use
P/N 11440-001 Rev. 3
Page 2-1
User Guide
2
Preparing the AutoPulse for Use
The AutoPulse System is delivered fully assembled, except for the LifeBand and the AutoPulse Battery.
2.1
LifeBand Load-distributing Band (LDB)
Note:
Do not cut the LifeBand before removing it from the AutoPulse. Cutting the LifeBand may
cause the AutoPulse to report a Fault and will require specific steps to clear that Fault.
2.1.1
Installing the LifeBand
1.
Power off the device.
2.
Place the AutoPulse Platform with the patient surface facing down on a smooth, flat surface.
Note:
The driveshaft should be oriented so that the slot faces directly upward.
3.
Insert the head end of the LifeBand band clip into the driveshaft slot. The correct direction is
towards the LifeBand plate alignment arrow seen on the Platform (see Figure 2-1).
Figure 2-1
Sliding the LifeBand Band Clip into the Driveshaft Slot
4.
Once the head end of the LifeBand band clip is positioned into the slot, press the tail end of
the band clip into the slot of the guide plate until the band clip is fully seated in the driveshaft. You
should feel it lock into place.
1
Cover Plate
LifeBand plate alignment arrow
Band Clip
Tail End
Head End
Slotted Drive Shaft
Guide Plate
1
2
2