3 manual calibration, 3 manual calibration -4, Operating procedures – Yokogawa DAQWORX User Manual
Page 177

8-4
IM WX102-01E
8.3
Manual Calibration
Operating Procedures
1
Click the connect button on the calibration screen. The DA100, DC100, or DR
system that is connected is displayed.
2
Clicking the module to be calibrated displays a calibration dialog box.
3
Click the Manual tab to display the manual calibration screen.
Zero current value
Zero calibration value
Span current value
Span calibration value
Displays input range
Displays input range
Start button
Range
Select all
ranges
Manual tab
4
Clicking the range to be calibrated shows the selected range in red. To calibrate
all ranges click All.
5
Set the calibration values for Zero and Span, individually. Clicking the Zero or
Span display area shows the setting range.
6
Clicking the Start button displays a message for writing the calibration value.
7
Clicking the OK button displays a final message for fixing the calibration value.
8
Clicking the OK button displays a message notifying that the calibration value has
been fixed.
9
Clicking the OK button terminates the calibration. To calibrate another module,
click the Close button and repeat from step 2. OK is displayed for the modules
that have completed the calibration.