Aquametrix 2200C Conductivity Controller User Manual User Manual
Page 11

4.7.1
The controls which are frequently used in the normal operation of the instrument are
all accessible on the control panel. Some switches, which are infrequently used,
are located on the back of the main circuit board.
4.7.2
The DIP switches are scanned only on power-up and every time the instrument is
taken out of the menu mode. Therefore, after changes to the DIP switch settings,
you must bring up the menu by pressing CALL, and then return the instrument to
the online mode by pressing RUN, in order for the instrument to scan the new DIP
values.
4.7.3
The following table describes the use of the 16 DIP switches:
Bank S1
Dip Switch Description
ON
OFF
1
Selects temperature unit
°C
°F
2
Password Feature
Disabled
Enabled
3
Fail-Safe mode for Relay A
Enabled
Disabled
4
Direction of Control for Relay A
Falling
Rising
5
Alarm Relay to be activated for memory loss
Disabled
Enabled
6
Fail-Safe mode for Alarm Relay
Enabled
Disabled
7
Auto return from menu
Enabled
Disabled
8
0-5V / 0-1mA Select
Conductivity Temperature
Bank S2
Dip Switch Description
ON
OFF
1
Cycling Control for Relay A
Enabled
Disabled
2
Reserved
3
Reserved
4
Reserved
5
Reserved
6
Reserved
7
Reserved
8
Reserved
Note: Items in “BOLD” are the factory settings that the unit will be shipped with.
4.8 Output Hold