Cal settings menu – DeFelsko PosiTector UTG v.3.0 Quick Guide User Manual

Page 7

Advertising
background image

Cal Settings Menu

Cal Settings Menu

The PosiTector UTG is factory calibrated. But in order for it to take

accurate thickness measurements of a particular material it must

be set to the correct sound velocity for that material. Be aware that

material composition (and thus its sound velocity) can vary from

stated tables and even between lots from a manufacturer.

Adjustment to a sample of known thickness of the material to be

measured ensures that the Gage is adjusted as close as possible

to the sound velocity of that specific material. Samples should be

flat, smooth and as thick as the maximum expected thickness of

the piece to be tested.
The PosiTector UTG allows four simple adjustment choices. All

four methods are based on the simple premise of adjusting the

sound velocity. The first three adjustment methods make 1-point

calibration adjustments to optimize the linearity of the PosiTector

UTG over small ranges. The fourth method makes a 2-point

calibration adjustment to allow for greater accuracy over a large

range.

Zero

PosiTector UTG probes can be "zeroed" to compensate for tem-

perature and wear. This action is not required when operating in

multiple-echo mode (UTG M probes only) except when the instru-

ment is new and after a Reset (pg. 8). However it ensures best

accuracy when operating in SE (single-echo) mode. Therefore it is

a good idea to "zero" the probe before each work shift and when

the temperature of the part changes. During zeroing, it does not

matter what the current velocity setting is.

1. Make sure the instrument is on and the probe is wiped clean.
2. Apply a single drop of couplant onto the zero plate located on

the underside of the probe connector. DO NOT apply couplant

directly onto the probe face.

3. Select the Zero menu option and follow the on-screen prompts.

Calibration Adjustment

6

Advertising