Caution: Fresh Air Calibration should be performed in an environment of fresh air without any

influence of other gases. The calibration is performed on the assumption that the concentration

of Oxygen is 20.9%, the Inflammable Gas is 0% LEL and the concentration of Toxic Gas is

0ppm in the fresh air. Accordingly, it is not recommended to perform Fresh Air Calibration in

a closed space, and calibration should not be performed where hazardous gases can be

inhaled by operators.

Single Gas Calibration

To enter Calibration Mode, press the Arrow key and Power key simultaneously for 3 seconds.
CAL will be displayed and the

icon will blink. Press the Arrow key again for Single Gas

Calibration. The icon will blink. Press the Power key for 5 seconds to start Single Gas

Calibration. Do not to proceed with calibration without connecting the single gas.

60secs count down

Concentration of

Disconnect from

Single Gas

Single Gas

When Single Gas Calibration starts, a 60 second countdown will be shown while the

Single Gas Calibration occurs. If the calibration is successful, the concentration value of
the gas connected at the moment will be indicated and the

icon will blink 3 times.

Afterwards, it will indicate the concentration value measured at the moment when the

Single Gas is disconnected. (Refer to the following table for correct gas concentration values)
If the calibration fails, the

icon will blink continuously. Re-calibrate and if the calibration

fails again, contact Detcon to replace the sensor.

Hold 5 secs

Calibration
Success

Calibration
Fail

60secs count down