Detcon 1000_H2S User Manual

Page 11

Advertising
background image

Model 1000 H2S Echem

Model 1000 H2S Echem Instruction Manual

Rev. 2.8

Page 7 of 30

AIR

FLOW

INLET

SAMPLE

SAMPLE

FLOW

10 PSIG (CONSTANT)

SAMPLE PRESSURE

INLET

AIR

AIR

MOD EL NO.

SERIAL NO.

INPUT VOLTAGE

F REQU ENCY

CURRENT L OAD

AREA CL ASSIFICATION

det con inc.

3200 A-1 Research Forest Dr.

The Woodlands, TX 77381

www.det con. com

PART NO.

Sample Inlet Port

Pressure Gauge

Sample Bypass and Liquid
Rejection Exhaust Vent.

NOTE:

These may be

combined into a

single vent

Sample Calibrate

H

2

S Vent

Cal Gas

Figure 8 Port Identification

2. Whenever possible, the Sample Bypass flow control valve of the Genie membrane filter should be used to

minimize the sample lag time between the sample tap and the analyzer location. It can also be used as a
means to exhaust condensed liquids in the sample line away from the Genie filter and prevent a “loss of
flow” condition. Set a flow of 100-200cc/min. and vent to a safe area using ¼” O.D. tubing.

3. Connect to the Over-pressure Relief valve and vent to a safe area. The pressure relief valve is set at

Detcon to open at 15-20psig. For convenience sake, the sample bypass and over-pressure relief can be
vented together
.

4. Install a length of ¼" OD stainless steel tubing from the vent port to an area deemed safe for venting as

shown below.

Venting pressure should be in the range of 0±1psig and ambient pressure is highly

preferred.

5. When the optional Acetic Acid Bubbler apparatus is used, prior to applying power and applying gas flow

the bubbler should be filled to the maximum fill level with a 5% acetic acid solution. (Standard white
vinegar is also acceptable).