B-2

SERIES EL 070110

B.2

NUMERICAL VALUES AND DATA CONVENTIONS

The following conventions are used throughout this Appendix.

< >

Items in these brackets are parameter abbreviations. Parameters may be numbers or text

<NR1>

Simple numbers (without a decimal point) from 0 - 255. Examples: 156, 0 or 1

<NR2 > Numeric Values with or without a decimal point. Examples: 0.0156, 15.6, 156 or 1.56E+3 (scientific

notation: digits with decimal point and exponent)

<txt>

Text data indicating. For example, operating modes: CURRent, VOLTage, POWer, RESistance,
CONDuctance, SHORt

<bool>

Boolean Data. Examples: 0, 1, OFF or ON. NOTE: If a command expecting a boolean argument is
sent with no argument, the argument is assumed to be 0 (e.g., INPut and INPut 0 are identical).

[ ]

Optional keywords are shown in square brackets. For example [SOURce:]INPut:[STATe] <bool>
means SOURce and STATe: may be omitted thus the command is reduced to INP <bool>.

B.3

CALibrate:IPADdress COMMAND

CAL:IPAD

Syntax:

Short Form: CAL:IPAD <a.b.c.d>
Long Form: CALibrate:IPADdress <a.b.c.d>
Where: <a, b, c, d> = numbers from 0 to 255 comprising the IP Address

Description: Sets the IP address to be used during LAN communication.

B.4

CALibrate:IPADdress? QUERY

CAL:IPAD?

Syntax:

Short Form: CAL:IPAD?

Long Form: CALibrate:IPADdress?

Returns: <a.b.c.d> = numbers from 0 to 255 comprising the IP Address

Description: Returns the IP Address used during LAN communication.

B.5

CALibrate:SUBNetmask COMMAND

CAL:SUBN

Syntax:

Short Form: CAL:SUBN <a.b.c.d>
Long Form: CALibrate:SUBNetmask <a.b.c.d>
Where: <a, b, c, d> = numbers from 0 to 255 comprising the Subnet Mask

Description: Sets the Subnet Mask to be used during LAN communication.

B.6

CALibrate:SUBNetmask? QUERY

CAL:SUBN?

Syntax:

Short Form: CAL:SUBN?

Long Form: CALibrate:SUBNetmask?

Returns: <a.b.c.d> = numbers from 0 to 255 comprising the Subnet Mask

Description: Returns the Subnet Mask used during LAN communication.

SYST:FAN, ?

B.107, B.108

SYST:RANG:VOLT, ?

B.117, B.118

SYST:PFM, ?

B.109, B.110

SYST:REP, ?

B.119, B.120

SYST:PFM:LIST?

B.111

SYST:SERN?

B.121

SYST:PFM:STAT?

B.112

SYST:SET?

B.122

SYST:RANG:CURR, ?

B.113, B.114

SYST:VERS?

B.123

SYST:RANG:POW, ?

B.115, B.116

TABLE B-1. SCPI SUBSYSTEM COMMAND/QUERY INDEX (CONTINUED)

COMMAND

PAR.

COMMAND

PAR.

Advertising