7 maintenance and fault rectification, 1 maintenance, 2 fault rectification – KROHNE OPTISOUND 3010C 2wire EN User Manual

Page 34: 7 maintenance and fault rectification

Advertising
background image

7 Maintenance and fault rectification

7.1 Maintenance

When used as directed in normal operation, OPTISOUND

3010 C is completely maintenance-free.

7.2 Fault rectification

OPTISOUND 3010 C offers maximum reliability. Nevertheless

faults can occur during operation. These may be caused by the

following, e.g.:

l

Sensor

l

Process

l

Power supply

l

Signal processing.

The first measures to be taken are to check the output signal

and evaluate fault messages via the indicating/adjustment

module. The procedure is described below.

Connect a hand-held multimeter with a suitable measuring

range acc. to the wiring plan.

?

4 … 20 mA signal not stable

l

level fluctuations

à

set integration time via the indicating/adjustment

module

?

4 … 20 mA signal missing

l

incorrect connection to power supply

à

Check connection acc. to chapter "Connection proce-

dure" and correct, if necessary, acc. to chapter "Wiring

plans"

l

no power supply

à

check cables for line break, repair, if necessary

l

supply voltage too low or load resistance too high

à

check and adapt, if necessary

Causes of malfunction

Fault rectification

Checking the 4 … 20 mA sig-

nal

34

OPTISOUND 3010 C - 4 … 20 mA/HART

Maintenance and fault rectification

30503

-EN
-050914

Advertising