KROHNE TTP 400 EN User Manual

Page 7

Advertising
background image

Installation and Operating Instructions TTP x00

7

3

Fault diagnosis and corrective action

In the event of a fault or maloperation, please go through the various faults listed in the table. Do
not attempt to disassemble the device.
Should you fail to locate the fault, please contact our Technical Service.

Fault Cause Action/elimination
With Pt100 connection
Incorrect display

Terminal assignment is
incorrect

Check terminal assignment

Cable break

Measure continuity

Error display

Short-circuit Check

wiring

With integrated signal converter

Supply voltage at module
< 8 V

Load impedance too high,
voltage too low

Incorrect current signal

Incorrect measuring range

Correct the programming

Cable break

Check continuity of cables

No current signal

Supply polarity incorrect

Reverse terminals 1, 2

Current signal > 23 mA

Short-circuit

Check wiring



Advertising
This manual is related to the following products: