Meaning of General Status Messages:

Status Message

Meaning

Field Device Malfunction

Set if any „Critical Error“ has been reported.
Refer to Transmitter Specific Status messages.

Configuration Changed

Set whenever any configuration parameter is changed.

Primary Variable Analog Output
Fixed

Set when:

Device is set in Multidrop Mode

(Polling Address

0)

Fixed Current Mode is entered

(during Loop Test and DAC Trim)

Analog Output Saturated

During normal operation the maximum value for analog output
is 20.4mA because of the „Namur Failure Signal“ (I>20.8mA).
While Process Value (PV) will work up to the Sensor Limits
the analog output and its digital value is saturated at 20.4mA
and the flag is set.

Primary Variable Out Of Limits

Is set whenever the Process Value (PV) exceeds the Sensor
Limits.

Meaning of Transmitter Specific Status Messages:

Status Message

Meaning

Critical Error

Error: Summary of „Initialization Failed“ and „Internal Checksum
Error“. If set the measurement is invalid.
HART-Classes: Hardware, State, Uncorrectable, Summary.

Initialization Failed

Error: If set the measurement is invalid.
HART-Classes: Hardware, State, Uncorrectable, Detail.

No Linearization Data

Warning: If set the measured value is possibly incorrect.
HART-Classes: Software, State, Uncorrectable.

Internal Checksum Error

Error: If set the measurement is invalid.
HART-Classes: Software, Event, Selfcorrecting/Uncorrectable,
Detail.

Internal Timeout

Informational: Self correcting error.
HART-Classes: Software, Event, Selfcorrecting.

ADC zero scale not calibrated

Warning: If set the measured value is possibly incorrect.
HART-Classes: Hardware, State, Uncorrectable.

ADC full scale not calibrated

Warning: If set the measured value is possibly incorrect.
HART-Classes: Hardware, State, Uncorrectable.

No TC correction data

Warning: If set the measured value is possibly incorrect.
HART-Classes: Software, State, Uncorrectable.