Technical data – KROHNE OPTIFLUX 1000 EN User Manual

Page 26

Advertising
background image

6

TECHNICAL DATA

26

OPTIFLUX 1000

www.krohne.com

04/2014 - 4000846303 - HB OPTIFLUX 1000 R03 en

Operating conditions

Temperature

Temperature

Temperature

Temperature

Process temperature

-25...+120°C / -13...+248°F

Ambient temperature

-25…+65°C / -13…+149°F

Protect electronics against self-heating at ambient temperatures above

+55° C / +131°F

Storage temperature

-50…+70°C / -58…+158°F

Measurement range

Measurement range

Measurement range

Measurement range

-12...+12 m/s / -40...+40 ft/s

Pressure

Pressure

Pressure

Pressure

Ambient pressure

Atmospheric

Operating pressure

Up to 16 bar / 230 psi

Vacuum load

0 mbar / psi absolute

Pressure loss

Negligible

Pressure ranges for

secondary containment

Pressure resistant up to 40 bar / 580 psi

Burst pressure up to approx. 160 bar / 2320 psi

Chemical properties

Chemical properties

Chemical properties

Chemical properties

Physical condition

Electrically conductive liquids

Electrical conductivity

Standard: ≥ 5 μS/cm

Demineralized water: ≥ 20 μS/cm

Permissible gas content

(volume)

IFC 050: ≤ 3%

IFC 100: ≤ 3%

IFC 300: ≤ 5%

Permissible solid content

(volume)

IFC 050: ≤ 10%

IFC 100: ≤ 10%

IFC 300: ≤ 70%

Installation conditions

Installation

Assure that the flow sensor is always fully filled.

For detailed information refer to

Installation

on page 12

Flow direction

Forward and reverse

Arrow on flow sensor indicates positive flow direction.

Inlet run

≥ 5 DN

Outlet run

≥ 2 DN

Dimensions and weights

For detailed information refer to

Dimensions and weights

on page 30