Installation – KROHNE OPTIMASS 6000 EN User Manual

Page 24

Advertising
background image

3

INSTALLATION

24

OPTIMASS 6000

www.krohne.com

07/2014 - 4001894003 - MA OPTIMASS 6000 R03 en

3.4.14 Zero calibration

The procedure for zero calibration is contained in the converter handbook. However, the
following information should be considered when installing the meter.

3.4.15 Sunshades

The meter MUST be protected from strong sunlight.

Zero calibration

1 Where the meter has been installed vertically, install shut-off valves either side of the meter to assist with zero cali-

bration.

2 If the process flow cannot be stopped, install a bypass section for zero calibration.

1 Horizontal installation
2 Vertical installation

MA_OPTIMASS_6000_en_140717_4001894003_R03.book Page 24 Thursday, July 17, 2014 3:27 PM

Advertising