5.10 Entry Code


Fct. 3.4 ENTRY CODE
Press key

.


Selection
NO

(no code, press key

to enter setting mode)

YES (enter setting mode with key

and Code 1:

→ → → ↵ ↵ ↵ ↑ ↑ ↑)

Select with keys

and .

Press key

to return to Fct. 3.4 ENTRY CODE.



5.11 Primary head

Fct. 3.2 FLOWMETER
Press key

.

DIAMETER = Set the meter size
(see instrument nameplate) press key

.

Select size from table of meter sizes:
BATCHFLUX IFM 5015 K :

DN 2,5 - 40

mm equivalent to

1

/

10

- 1

1

/

2

inches


Select with keys

and .

Transfer to subfunction “FULL SCALE“
with key

.

FULL SCALE = Set the full-scale range,
press key

.

Set as described in Sect. 5.1.

Transfer to subfunction “GKL VALUE“
with key

.


Note: if “VALUE P“ is displayed after
pressing key

↵ .

PULSE/VOL. has been set under Fct. 1.6
PULS.OUTP. P, subfunction “SELECT P“.
Because the full-scale range Q

100%

has been

changed, the output frequency (F) of the pulse
output is over- or undershot:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Sect. 5.7
Pulse output P, Fct. 1.6.


GKL VALUE = Set the primary
constant GK,
press key

.

1.0000 - 9.9999

(note information on

instrument nameplate, do not change
setting !)


Change flashing digit (cursor) with keys

and

. Shift cursor 1 place to right or left with keys

and .
Transfer to subfunction “FLOW DIR.“ with
key

.


FIELD FREQ. = Magnetic field
frequency,
press key

.

1/2

(1/2, 1/6 or 1/18 or 50 Hz

or 60 Hz,

1/6

see instrument nameplate,

1/18 do

not change setting !)

The magnetic field frequency is determined by
the program.

Select with keys

and .

Transfer to subfunction “FLOW DIR.“ with
key

.

FLOW DIR. = Set the direction of flow,
press key

.

+ DIR.

- DIR.

(for identification of flow

direction see “+“ arrow on primary head, in
F/R mode this identifies the “positive“ flow
direction)


Select with keys

and .

Press key

to return to Fct. 3.2

FLOWMETER.

Zero check: see Fct. 3.3 and Sect. 7.1.

Please refer to Sect. 3.2
“Factory settings“