KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 62

Advertising
background image

62

Driving the control input

5.10

Setting the system control

Some applications for the MFM 4085 may require measurement to be suspended at particular
times, such as during steam cleaning. The system control functions allow the converter to
automatically detect user-defined conditions and then to act accordingly.

Selectable conditions (Fct. 3.7.2.) are :

Density out of range
Temperature out of range

The ranges for these conditions are set by Fct. 3.7.3. and 3.7.4. (These range checks are
subject to hysteresis in the same way as the process alarm, see Sect. 5.7).

When such a condition is met, the converter takes one of the following actions :

1. Flow reading forced to zero, totaliser stopped, any output registering flow will go to zero.

2. Flow reading forced to zero as above, but totaliser is cleared to zero when measurement

restarts.

3. Disable outputs. All outputs, current, frequency and alarm are forced to their zero/inactive

state.


Example:

A process has to be steam cleaned regularly. The user has the pulse output set to total mass,
but he does not want pulses sent to his instrumentation during cleaning. However, he still
requires a temperature reading from the current output. The nominal density of his process fluid
is 1.2g/cm3.