The procedure for performing a two point density calibration is as follows.

Set the Process

Density for the First

Point

Set the Product type

(Holding Reg. 1028)

Product Type = 3 ?

YES

Set the Density Value

(Holding Reg. 3093)

Start Two Point

Calibration

(Set Output Reg. 1012)

YES

NO

Error During Density

Calibration (See notes)

Set the Process

Density for the

Second Point

Set the Product type

(Holding Reg. 1028)

Density Calibration

Status = 1 ?

(Input Reg 1007)

NO