KYORITSU 2433 User Manual
Kew snap, Series
INSTRUCTION MANUAL
MODEL 2432/2433
KEW SNAP
Series
LEAKAGE CURRENT TESTER
MODEL 2432
MODEL 2433
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
1. SAFETY WARNINGS
This instrument has been designed and tested according to IEC
Publication 61010: Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety
rules which must be observed by the user to ensure safe operation
of the instrument and to retain it in safe condition. Therefore, read
through these operating instructions before starting using the
instrument.
WARNING
●Read through and understand instructions contained in this
manual before starting using the instrument.
●Save and keep the manual handy to enable quick reference
whenever necessary.
●Be sure to use the instrument only in its intended applications
and to follow measurement procedures described in the
manual.
●Be sure to understand and follow all safety instructions
contained in the manual.
Not following the above instructions may cause injury, instrument
damage and/or damage to equipment under test.
The symbol
indicated on the instrument means that the user
must refer to related parts of the manual for safe operation of the
instrument. Be sure to carefully read the instructions following
each
symbol in this manual.
DANGER is reserved for conditions and actions that are likely
to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can
cause serious or fatal Injury.
CAUTION is reserved for conditions and actions that can cause
minor injury or Instrument damage.
DANGER
●Never make measurement on a circuit having potential of
300VAC or greater.
●Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the instrument may
cause sparking, which leads to an explosion.
●The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible
shorting where the equipment under test has exposed metal
parts.
●Never attempt to use the instrument if its surface or your hand is
wet.
●D o n o t e x c e e d t h e m a x i m u m a l l o w a b l e i n p u t o f a n y
measurement range.
●Never open the battery compartment cover when making
measurement.
WARNING
●Never attempt to make any measurement, if any abnormal
conditions are noted, such as broken case, cracked test leads
and exposed metal parts.
●Do not turn the range selector switch while the test leads are
connected to the circuit under test.
●Do not install substitute parts or make any modification to the
instrument. Return the instrument to Kyoritsu or your distributor
for repair or re-calibration.
●Do not try to replace the batteries if the surface of the
instrument is wet.
●Always switch off the instrument before opening the battery
compartment cover for battery replacement.
CAUTION
●Make sure that the range selector switch is set to an appropriate
position before making measurement.
●Do not expose the instrument to the direct sun, extreme
temperatures or dew fall.
●Be sure to set the range selector switch to the "OFF" position
after use. When the instrument will not be in use for a long
period of time, place it in storage after removing the batteries.
●Use a damp cloth and detergent for cleaning the instrument. Do
not use abrasives or solvents.
2. FEATURES
●Digital clamp tester for AC leakage measurement.
●Least affected by external magnetic field, providing wide
measuring range from very small to large currents.
●Designed to safety standard IEC 61010-2-032: over-voltage
category CAT.Ⅲ, 300V and pollution degree 2.
●Tear drop shaped jaws for ease of use in crowded cable areas
and other tight places.
●Data hold function to allow for easy readings in dimly lit or hard-to-
reach locations.
●Provides filtering function to remove high frequency generated by
such equipment as inverters.
●Peak hold function to allow for measurement of current variation
as short as 10msec.
●Auto-power-off function prevents unnecessary power consumption
●Dynamic range of 4000 counts full scale.
●Large easy-to-read LCD display with letter height of 13mm.
●Operation confirming beeps.
●Insulation barrier at the tip of transformer jaws for improved safety.
3. SPECIFICATIONS
Measuring ranges and accuracy
Operating System:
Sequential comparison
Display:
Lliquid crystal display with maximum reading
of 3999
Low battery warning:
"BATT" mark appears on the display
Overrange Indication:
"OL" appears on the display when upper limit
of measuring range is exceeded
Response Time:
Approx. 2 seconds
Sample Rate:
Approx. 2.5 times per second
Accuracy-insured
23℃±5℃, relative humidity 85% or less
Temperature and
(without condensation)
Humidity Ranges:
Operating Temperature 0-40℃, relative humidity 85% or less
and Humidity Ranges:
(without condensation)
Storage Temperature
-20-60℃, relative humidity 85% or less
and Humidity Ranges:
(without condensation)
Power Source:
Two 1.5V R03 (UM-4) batteries
Current Consumption:
Approx. 13mA
Measurement Time:
Approx. 40 hours
Auto-power-off Function:Turns power off about 10 minutes after the
last switch operation
Safety Standard:
IEC 61010-2-032
over-voltage CAT.Ⅲ 300V, pollution degree 2
IEC61326 (EMC)
Overload Protection:
M-2432: 120AAC max. for 10 seconds
M-2433: 480AAC max. for 10 seconds
Withstand Voltage:
3700VACrms (50/60Hz) for 1 minute between
metal part of transformer jaws and housing
case (except transformer jaw case)
Insulation Resistance:
10MΩ or greater at 1000V between metal
part of transformer jaws and housing case
(except transformer jaw case)
Conductor Size:
Approx. 40mm in diameter max.
Dimensions:
185(L)×81(W)×32(D)mm
Weight:
M-2432: Approx. 290g including batteries
M-2433: Approx. 270g including batteries
Accessories:
Two R03 (UM-4) batteries
Carrying case Model 9052
Instruction manual
Optional Accessories:
Multi-Tran Model 8004 and 8008
Model
M-2432
M-2433
Range
4mA
40mA
100A
40mA
400mA
400A
Measuring Range
0∼3.999mA
0∼39.99mA
0∼100.0A
0∼39.99mA
0∼399.9mA
0∼399.9A
Accuracy (Frequency range)
±1.0%rdg±5dgt
(50/60Hz)
±2.5%rdg±10dgt (20∼1kHz)
0∼ 80.0A
±1.0%rdg±5dgt
(50/60Hz)
±2.5%rdg±10dgt (40∼1kHz)
∼100.0A
±5.0%rdg
(50/60Hz)
±10.0%rdg
(40∼1kHz)
±1.0%rdg±5dgt
(50/60Hz)
±2.5%rdg±10dgt (20∼1kHz)
0∼350.0A
±1.0%rdg±5dgt
(50/60Hz)
±2.5%rdg±10dgt (40∼1kHz)
∼399.9A
±2.0%rdg
(50/60Hz)
±5.0%rdg
(40∼1kHz)
When measuring current which pulse element is superposed, differences
of the indicated value may be caused between ranges, if the peak value
exceeds the measurement range to a large extent. In this case, the
reading at the bigger range should be taken as a right value.
4. INSTRUMENT LAYOUT
5. PREPARATIONS FOR MEASUREMENT
5-1 Checking Battery Voltage
Set the Range Selector Switch to any position other than the OFF
position. If the marks on the display is clearly visible without "BATT"
mark showing, battery voltage is OK. If the display blanks or "BATT"
is indicated, replace the batteries according to section 8: Battery
Replacement.
NOTE
When the instrument is left powered on, the auto-power-off
function automatically shut the power off; The display blanks even
if the Range Selector Switch is set to a position other than the
OFF position in this state. To power on the instrument, turn the
Range Selector Switch or press the Data Hold Button. If the
display still blanks, the batteries are completely exhausted.
Replace the batteries.
5-2 Checking Switch Setting
Make sure that the Range Selector Switch is set to the appropriate
range.
A l s o m a k e s u r e t h a t d a t a h o l d f u n c t i o n i s n o t e n a b l e d . I f
inappropriate range is selected, desired measurement cannot be
made.
Transform Jaws
Trigger
Range Selector Switch
Hand Strap
Display
Frequency Selector Button
Peak Hold Button
Data Hold Button
●LCD
Low Battery Warning
Frequency Response :
Wide
Frequency Response :
50/60Hz
Data Hold Indication
Peak Hold Indication
Function
6. OPERATING INSTRUCTIONS
6-1 Current Measurement
DANGER
●In order to avoid possible shock hazard, never make measurement
on circuits having a potential of 300VAC or greater.
●The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible
shorting where the equipment under test has exposed metal parts.
●Never make measurement with the battery compartment cover
removed.
●When measuring current is 300A or more ( 400Hz or more ), be sure to
stop measurement within 5 minutes. Otherwise, transformer jaws may
heat to cause a fire or deformation of molded parts, which will degrade
insulation.
CAUTION
●Take sufficient care to not to apply shock, vibration or
excessive force to the jaw tips. Otherwise, precisely adjusted
Transformer Jaw tips will be damaged.
●When a foreign substance is stuck in the jaw tips or they
cannot properly engage, the transformer jaws do not fully
close. In such a case, do not release the jaw trigger abruptly or
attempt to close the transformer jaws by applying external
force. Make sure that the jaws close by themselves after
removing the foreign substance or making them free to move.
●The maximum size of a conductor to be tested is 40mm in
diameter. Accurate measurement cannot be made on a
conductor larger than this, because the transformer jaws
cannot fully close.
●When measuring large current, the transformer jaws may buzz.
This has no effect on the instrument's performance or safety.
(1)Set the Range Selector Switch to the desired position. Current
to measure should be within the selected measuring range.
(2)Normal measurement (See Fig.1):
Press the jaw trigger to open the transformer jaws and close
them over one conductor only. Measured current value is shown
on the display. Earth leakage current or small current that flows
through a grounded wire can also be measured by this method.
(3)Measuring out of balance leakage current (See Fig. 2):
Clamp onto all conductors except a grounded wire. Measured
current value is shown on the display.
Fig. 1 Normal measurement
Correct
Wrong
Load
Load
6-2 How to Use Frequency Selector Button
When high frequencies from such equipment as inverters are
present in the circuit under test, the instrument measures AC
current of not only 50Hz or 60Hz of fundamental frequency but also
of these high frequencies and harmonics.
To eliminate the effect of such high frequency noise and measure
AC current of 50Hz or 60Hz fundamental frequency, a "high-cut"
filter circuit in incorporated into the instrument which works when
"50/60Hz" frequency response is selected with the Frequency
Selector Button. Cut-off frequency of the "high-cut" filter is about
160Hz with attenuation characteristic of approx. -24dB/octave.
When the Frequency Selector Button is pressed, "50/60Hz" mark is
shown on the left side of the display. When the Frequency Selector
Button is pressed again, fequency response is switched to WIDE
with "WIDE" mark shown on the display.
Note:
Characteristic of -24dB/octiave means that signal magnitude
declines to about one sixteenth of that at the initial frequency when
frequency doubles. Model 2432 and Model 2433 have the following
two settings for the Frequency Selector Button.
WIDE (20Hz-): Permits measurement of currents of fundamental
frequencies as well as currents of high frequencies
generated by such equipment as inverters
50/60Hz (20-approx.160Hz): Filters out high frequency currents and
measures current of fundamental
frequency only
Fig. 2 Measuring out of balance leakage current
3-phase 3-wire system
(In 4-wire system with neutral,
clamp onto all 4 wires)
Single-phase 2-wire system
(in 3-wire system with neutral,
clamp onto all 3 wires)
Fig.3 Model 2432/2433 Frequency Characteristic
Output characteristic are shown in Fig.3.
;;;;
;;;;
;;;;
;;;;
;;;;
;;;;
;;;;
100%
50%
20%
10%
5%
5k
1k 2k
500
100 200
50
20
10
ー30
ー20
ー10
0
Attenuation〔dB〕
10
Frequency〔Hz〕
Percentage
Typical characteristic:
4/40/400mA Range
Typical characteristic:
“WIDE”frequency response
Typical characteristic:
“50/60Hz”frequency response
high-cut filter -24dB/oct
Typical characteristic:
100/400A Range
Load
Load
Recently there has been increased use of power through inverters,
switching regulators, etc. When the high frequency noise from such
appliances leaks or flows into the ground through capacitors not
filtering completely, the earth leakage breaker may trip even though
there is no "actual" leakage. In such a case, the instrument do not
give leakage current reading if "50/60Hz" frequency response is
selected.
Take current readings with the 50/60Hz and WIDE frequency
responses respectively to make effective use of the Frequency
Selector Button.
6-3 Peak Current Measurement
(1)Set the Range Selector Switch to the desired position.(Current to
measure should not exceed the selected measuring range.)
(2)Select "WIDE"or "50/60Hz"with the Frequency Selector Button.
(3)With the transformer jaws clamped onto the conductor under
test, press the Peak Hold Button to set the interment to the peak
measurement mode.("P" is shown on the display.)
(4)The display reads 1/√2 of the peak current value. Therefore,an
rms reading is shown when current of a sinusoidal waveform is
measured.
(5)After peak measurement, press the Peak Hold Button to return to
the normal measurement mode.
Note: When leakage current is measured in the peak measurement
mode, the reading may change if the transformer jaws are
opened and closed. Please read the display with the conductor
under test clamped, otherwise, after fixing the display by using
the data hold function, please remove the instrument from the
conductor to be measured, and read the display. To measure
the peak current again, please release the data hold, and return
the instrument to the normal measurement mode once with the
Peak Hold Button, then set it in the peak measurement mode.
7. OTHER FUNCTIONS
7-1 Auto-Power-Off Function
This is a function to prevent the instrument from being left powered
on and conserve battery power. The instrument automatically turns
off about 10 minutes after the last switch or button operation. To
return to the normal mode, turn the Range Selector Switch to OFF,
then to the desired position.
Disabling Auto-Power-Off Function:
To disable the auto-power-off function, power on the instrument
with the Data Hold Button pressed. About 3 seconds after powering
on the instrument, "P.OFF" is shown on the display. To enable the
auto-power-off function, turn on the instrument without pressing the
Data Hold Button.
Note: The auto-power-off function is disabled in the peak
measurement mode.
7-2 Date Hold Function
This is a function to freeze the readings on the display. When the
Data Hold Button is pressed once, the current reading is held even
though current under test varies. "H" mark is shown on the upper
right corner of the display.
To exit the data hold mode, press the Data Hold Button again.
Note: When the auto-power-off function works while the instrument
is in the data hold mode, data hold is cancelled.
8. BATTERY REPLACEMENT
WARNING
In order to avoid possible shock hazard, always set the Range
Selector Switch to the OFF position before trying to replace the
batteries.
CAUTION
●Do not mix new and old batteries.
●Install batteries in the orientation as shown inside the battery
compartment, observing correct polarity.
When the battery voltage warning mark "BATT" is shown on the
top left corner of the LCD, replace the batteries. Note that the
display blanks and "BATT" mark is not shown if the batteries are
completely exhausted.
(1)Set the Range Selector Switch to "OFF."
(2)Loosen the battery-compartment-
cover-fixing screw on the lower
back of the instrument.
(3)Replace the batteries with two
new R03 (UM-4) 1.5V batteries.
(4)Put the battery compartment
cover back in place and tighten
the screw.
Note: For use for a long period of time,
use alkaline batteries.
9. OPTIONAL ACCESSORIES
Model 8004 and 8008 (Multi-Tran)
These models help Model 2432 or Model
2433 to measure current greater than
3000A or to make measurement on a
large bus-bar or conductor.
(1)Set the Range Selector Switch to
"100A" or "400A."
(2)As shown, open the jaws and close
them over the pickup coil of Model
8004 or Model 8008.
(3)Clamp on a conductor with Model
8004 or Model 8008.
(4)Take the reading and multiply it by
10.
Note: When used with Model 8008, Model 2432 can measure up to
1000A.
Model 8004 and Model 8008 cannot be used for leakage current
measurement. For detailed specifications, refer to the instruction
manual for Model 8004 or Model 8008.
M-2432
M-2433
M-8004
M-8008
Pickup Coil
Conductor
Transformer
jaws
10
:
1
M-8004
M-8008
Max. Conductor Size
60mm in diameter
100mm in diameter
Measuring
Range
0∼1000A
0∼3000A
Current Transformation
Ratio
10:1
10:1
Screw
Battery
Compartment
Cove
Batteries
92―1452B
01―07
DISTRIBUTOR
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.
No.5-20, Nakane 2― chome, Meguro-ku,
Tokyo, 152-0031 Japan
Phone:
(03)
3723― 0131 Fax:
(03)3723― 0152
URL:http://www.kew-ltd.co.jp
E-mail:[email protected]
Factories:Uwajima&Ehime
R
Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.