Dwyer HM35 User Manual

Page 16

Advertising
background image

15

Parameter
limits

Feedback data

Feedback data
type

Description

Remarks

0, 10, 25

<value>

FLOAT

Continuous transfer of
measured values

Continuous measurement
with 10 or 25 M/s.
Stops with Interval=0

--- <value>

FLOAT

Query

measured

pressure value

Individual value

--- <value>

FLOAT

Query

sensor

temperature

°C, individual value

mbar, bar,.... atm

---

---

Pressure unit input

---

<unit>

DISCRETE

Query pressure unit

E.g., mbar, bar,.... atm

yyyy: 2001 ...2099
mm: 1...12
dd: 1... xx

---

---

Input date

yyyy: year,
mm: month
dd: day

---

<yyyy>,<mm>,<dd> INTEGER

Query date

yyyy: year, mm: month
dd: day

0...23, 0...59, 0...59 ---

---

Input time

hh: hours, mm: minutes
ss: seconds

--- <hh>,<mm>,<ss>

INTEGER

Query

time

hh: hours, mm: minutes
ss: seconds

--- <Error_number>,

"<Error_description>
(;<Device-
dependent info>)"

INTEGER,
STRING

Query SCPI Error Queue STRING with ‚fixed text‘

and optional ‚free text’,
separated by a semicolon,
maximum 255 digits

---

<version>

FLOAT

SCPI query and firm-
ware version

e.g. '2001.0' , ‘FW:300’

ON, OFF

---

---

Beeper enable / disable

---

<value>

INTEGER

Query battery state

Range 0...100, value in %

--- <range>

STRING

Query sensor
measurement range
(in mbar)

e.g. "1,000 mbar“

--- <tolerance>

STRING

Query

sensor tolerance

e.g. '0.05 %FS'

---

<type, MOD, S/N>

STRING Query

instrument

identification

e.g. "HM3500DLH200,
MOD00A,1234567“

ON, OFF

---

---

Set filter for display

--- --- ---

Zero

measure

pressure

value (ZERO)


Advertising