Specification – Dwyer UFC User Manual

Page 54

Advertising
background image

8: Specification

50

TRANSDUCER SETS

Standard:

Temperature Range -4°F to +275°F (-20°C to +135°C).
'A-ST' (standard) 0.5 to 4.5 inches (13 to 115 mm) pipe OD (2 MHz).
'B-ST' (standard) 2 to 79 inches (50 to 2000 mm) pipe OD (1 MHz).
Protection: IP54.

Optional:

'D' 59 to 197 inch (1500 to 5000 mm) pipe OD (0.5 MHz), with an operating temperature range of
-4°F to +176°F (-20°C to +80°C).

LANGUAGES

Standard Supported Languages:

English, French, German, Italian, Spanish, Portuguese, Russian, Norwegian, Dutch, Swedish.

DATA LOGGER

Data Logged:

All measurement parameters and flow readings, and both Totalized values.

Log Capacity:

The internal SD card can store more than 4 years of data at the fastest logging rate.

Time Stamping:

All data points are date and time stamped.

Number of Sites:

1 non-editable site plus 1 editable site.

Logging Interval:

User programmable, 5 seconds to 1 hour.

Logged Data Transfer:

Logged data can be transferred to a USB memory stick.

OUTPUTS

Current Output:

No. Channels
Format:
Resolution:
Error Currents:
Isolation:
Maximum Load:

1
4 to 20 mA, 0 to 20 mA, 0 to 16 mA.
0.1% of full scale.
Any between 0 to 26 mA.
1000 V Opto-isolated from unit.
620 Ohms.

Pulse Output TTL:

Number Available:
Isolation:
Pulse Repetition Rate:
Pulse Width:
Max Current:
Max Voltage:

1 – Opto-isolated MOSFET relay.
1500 V opto isolated from unit.
User programmable from 1 to 250 pps.
User programmable from 2 ms to 500 ms.
500 mA.
48V.

Alarms:

Number of Channels Available:
Isolation:
Relay Contact Mode:

Alarm Functions:

Frequency output:
Max Current:
Max Voltage:

2 off opto-isolated MOSFET relay.
1500 V opto isolated from unit.
N/O when switching condition is False.
Closed when switching condition is True.
The two relays can be configured to operate when:

a predefined

MIN

imum or

MAX

imum flow rate is exceeded.

a specified

VOL

ume is measured.

a signal Error condition occurs.

manual alarm test.

pulse frequency.

Maximum 200 Hz.
500 mA.
48V.

Advertising