The scan distance function is accessed from the default screen. AUTO. CAL is
displayed at the base of the display screen during the scan distance operation,
indicating that you are working in scan distance mode.

To Set Scan Distances:

Clearing the Scan Distance Values
The AUTO function enables you to clear all saved scan distance values. (Refer to
Setting the Scan Distance Values, page 12, for a description of how to set the scan
distance values.)

To Clear Scan Distance Values:

Viewing Processor Information
You can view the version number and absolute address for each of the three
processors contained in Series UT, as follows:

Screen processor, indicated by S before the code
Main processor, indicated by h before the code
Co processor, indicated by C before the code

To View Processor Information:

Chapter 4: Series UT Open Channels

This chapter describes how to set flow measurement parameters for open channels
using Series UT models, and explains the flume/weir codes methodology used
when setting up flow measurements.

Selecting the Flow Measurement Settings
The PARSH.FLUM function in the main menu enables you to select one of the
preset flumes/weirs settings for flow measurements.

When setting flow measurement parameters, the flume/weir type value (X) is
entered first, followed by the letter (U) or (E) as an indication for American or
European open channel flow standard, and followed by the code value (YY) that
represents the appropriate flume/weir dimensions, in the following format: .
The default is European standard. The open channel types and codes are
described in Open Channels Flow Measurements, page 14.

If you wish to insert custom flume measurements, you must enter .001 for this
function. This entry will automatically initiate an additional menu function (Pr 1),
enabling manual insertion of custom flume values in an accordance table, as
described in Chapter 5, Additional Features.

Page 13

NOTES:
The reading of the actual target height may not be exact; for example, a target
height of 6 m may give a reading of 5.998.

Press/Action
From the default
screen, press

and

simultaneously.

Display

Explanation

Displayed temporarily while Series UT
searches for an interfering signal.

and

NOTES:
Do not press any key while the SEARCH message is displayed.

Press/Action

(Wait a few
seconds.)

Display

For example,

For example,

Explanation

Displays the depth of the interfering
signal.

Saves the interfering signal, then
searches again and displays the next
reading. Continue this process to save
up to eight interference readings.

Actual target height reading indicates
that there are no more interfering
signals.

Saves the true echo value and
completes the scan distance operation.

NOTES:
Pressing the NEXT button saves identified interferences. Pressing the ENT button
exits the function while saving the true echo value.

Press/Action Display

Explanation

Select AUTO function using Next
or Back buttons.

With the required option flashing.

NOTES:
If the indication at the base of the default display screen is lit, it might be a
sign of acoustic interferences in the application. It is likely that Series UT will
overcome these interferences (i.e. display a correct value), by using a different
transmission sequence.

Press/Action

In the main menu
display, press

and

simultaneously.

Press on the
far right digit.

Display

For example,

For example,

Explanation

Use to enter the value 000.003.

The version number of the screen
processor is displayed.

Use to scroll through the version number
and absolute address for each
processor.

Press at any time to return to the main
menu.

or

or

or

NOTES:
Refer to Chapter 3, Basic Setup, for an explanation of accessing and using the
Series UT main menu and function buttons.

or

Advertising
This manual is related to the following products: