Particulate Monitoring Systems

Installation & Operating Manual

Document No. 210-1003-M

Page 36

©2007

Modbus®/RTU Registers

Register Description Data

Address

0x Reference: Discrete output coils

Type

00001 Alarm relay 1 (1=ON, 0=OFF)

Bit

00002 Alarm relay 2 (1=ON, 0=OFF)

Bit

03073 Alarm status - alarm 1 (1=ON, 0=OFF)

Bit

03074 Alarm status - alarm 2 (1=ON, 0=OFF)

Bit

03247 Zero Check Running Status (1=Running, 0=Not Running)

Bit

03248 Span Check Running Status (1=Running, 0=Not Running)

Bit

03249 Full System Zero Check Running Status (1=Running, 0=Not Running)

Bit

03251 Zero Check Remote Activation (Set to 1 to Activate)

Bit

03252 Span Check Remote Activation (Set to 1 to Activate)

Bit

03254 Probe & Cable Check Status (1=Error, 0=OK)

Bit

03255 Zero Check Status (1=Error, 0=OK)

Bit

03256 Span Check Status (1=Error, 0=OK)

Bit

03257 Full System Zero Check Status (1=Error, 0=OK)

Bit

Description

4x Reference: Holding registers

40609 Process Variable 1 - particulate (pA)

Float

40611 Process Variable 2 - averaged particulate (pA)

Float

42309 Correlation Scaling Raw Minimum (pA)

Integer

40705 Correlation Scaling Raw Maximum (pA)

Integer

40641 Correlation Scaling Scaled Minimum (mg/m3 or gr/ft3)

Integer

40647 Correlation Scaling Scaled Maximum (mg/m3 or gr/ft3)

Integer

40721 Process Variable Units of Measure (0=pA, 1=mg/m3, 2=gr/ft3)

Integer

40737 Alarm 1 Level - (pA)

Float

40739 Alarm 2 Level - (pA)

Float

40769 Alarm 1 delay - (sec)

Integer

40770 Alarm 2 delay - (sec)

Integer

43363 Zero Check Delay Between Checks (Hours)

Float

43365 Span Check Delay Between Checks (Hours)

Float

43367 Full System Zero Check Delay Between Checks (Hours)

Float

43381 Probe & Cable Check Relay to Activate (0=None, 1=Relay 1, 2=Relay 2)

Integer

43382 Zero Check Relay to Activate (0=None, 1=Relay 1, 2=Relay 2)

Integer

43383 Span Check Relay to Activate (0=None, 1=Relay 1, 2=Relay 2)

Integer

43384 Full System Zero Check Relay to Activate (0=None, 1=Relay 1, 2=Relay 2)

Integer

43399 Zero Check Allowable Tolerance (pA)

Float

43404 Span Check Allowable Tolerance (pA)

Float

43409 Full System Zero Check Allowable Tolerance (pA)

Float

43401 Zero Check Settling Time (milliseconds)

Integer

43406 Span Check Settling Time (milliseconds)

Integer

42402 Result Measured from Last Zero Check (pA)

Float

42407 Result Measured from Last Span Check (pA)

Float

42412 Result Measured from Last Full System Zero Check (pA)

Float

Advertising
This manual is related to the following products: