AirSep MN166-1 B User Manual

Page 34

Advertising
background image

30

MN166-1

rev

B

FreeStyle 5 Specifications

Oxygen Concentration:*

1-5 pulse settings; equivalent to a continuous flow of 90% oxygen +5.5 / -3%

Pulse Dose:

Setting 1: 8.75ml

± 15%

Setting 2: 17.5ml

± 10%

Setting 3: 26.5ml

± 10%

Setting 4: 35.0ml

± 10%

Setting 5: 43.75ml

± 10%

Dimensions:

10.7 in. high x 6.6 in. wide x 4.4 in. deep (27.2 cm high x 16.8 cm wide x 11.2 cm
deep)

Weight:

6.2 lb (2.8 kg); 1.8 lb (.8 kg) optional AirBelt

Power:

AC Power Supply: Input: 100-240VAC, 50-60Hz, 2A | Output: 15V

DC Power Supply: Input: 12-24VDC, 15A max | Output: 15VDC, 8A max

Battery duration
(Rechargeable lithium
Battery)

5 LPM – 1 hour; 4 LPM – 1 hour, 15 minutes; 3 LPM – 1 hour, 30 minutes; 2 LPM –
2 hours; 1 LPM – 3 hours
Optional AirBelt when combined with the internal battery:
5 LPM – 2 hours, 30 minutes; 4 LPM – 3 hours; 3 LPM – 3 hours, 30 minutes;
2 LPM – 4 hours, 30 minutes; 1 LPM – 7

hours

Battery recharge time:

3

hours, 15 minutes; optional AirBelt:

3

hours

Initial startup time

2 minutes

Battery cycle life:

Approximately 300 cycles, then 80% capacity or below.

Audible alarms and pulse
visual indicators:

Start-up – audible and visual (GREEN indicator light);
Pulse flow – visual (GREEN indicator light);
Battery condition – battery level (GREEN indicator lights);
Battery Low – audible and visual (YELLOW alarm light);
Battery shutdown – audible and visual (YELLOW alarm light);
Cannula disconnect – audible and visual (YELLOW alarm light);
System overdraw – audible and visual (YELLOW alarm light);
General malfunction – audible and visual (RED alarm light).

**Temperature range:

Operational temperature: 41°F to 104°F (5°C to 40°C)
(Up to 95% RH (non-condensing)
Storage temperature:

-4°F to 140°F (-20°C to 60°C)

**Max Operational
Altitude:

up to 12,000 ft (3,657.6 m) (483mmHg) Higher altitudes may affect performance

* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)

**Operating outside of these specifications can limit the concentrators ability to meet Oxygen Concentration specifications at higher liter flow rates.