AirSep MN166-1 B User Manual
Page 34

30
MN166-1
rev
B
FreeStyle 5 Specifications
Oxygen Concentration:*
1-5 pulse settings; equivalent to a continuous flow of 90% oxygen +5.5 / -3%
Pulse Dose:
Setting 1: 8.75ml
± 15%
Setting 2: 17.5ml
± 10%
Setting 3: 26.5ml
± 10%
Setting 4: 35.0ml
± 10%
Setting 5: 43.75ml
± 10%
Dimensions:
10.7 in. high x 6.6 in. wide x 4.4 in. deep (27.2 cm high x 16.8 cm wide x 11.2 cm
deep)
Weight:
6.2 lb (2.8 kg); 1.8 lb (.8 kg) optional AirBelt
Power:
AC Power Supply: Input: 100-240VAC, 50-60Hz, 2A | Output: 15V
DC Power Supply: Input: 12-24VDC, 15A max | Output: 15VDC, 8A max
Battery duration
(Rechargeable lithium
Battery)
5 LPM – 1 hour; 4 LPM – 1 hour, 15 minutes; 3 LPM – 1 hour, 30 minutes; 2 LPM –
2 hours; 1 LPM – 3 hours
Optional AirBelt when combined with the internal battery:
5 LPM – 2 hours, 30 minutes; 4 LPM – 3 hours; 3 LPM – 3 hours, 30 minutes;
2 LPM – 4 hours, 30 minutes; 1 LPM – 7
hours
Battery recharge time:
3
hours, 15 minutes; optional AirBelt:
3
hours
Initial startup time
2 minutes
Battery cycle life:
Approximately 300 cycles, then 80% capacity or below.
Audible alarms and pulse
visual indicators:
Start-up – audible and visual (GREEN indicator light);
Pulse flow – visual (GREEN indicator light);
Battery condition – battery level (GREEN indicator lights);
Battery Low – audible and visual (YELLOW alarm light);
Battery shutdown – audible and visual (YELLOW alarm light);
Cannula disconnect – audible and visual (YELLOW alarm light);
System overdraw – audible and visual (YELLOW alarm light);
General malfunction – audible and visual (RED alarm light).
**Temperature range:
Operational temperature: 41°F to 104°F (5°C to 40°C)
(Up to 95% RH (non-condensing)
Storage temperature:
-4°F to 140°F (-20°C to 60°C)
**Max Operational
Altitude:
up to 12,000 ft (3,657.6 m) (483mmHg) Higher altitudes may affect performance
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
**Operating outside of these specifications can limit the concentrators ability to meet Oxygen Concentration specifications at higher liter flow rates.