MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual

Rev B

5.10 Measuring System Pressure

1. Remove the back cover.

2. Locate the brass fitting with the orange plug on the left side of the feed/waste

manifold.

3. Remove the plug.

4. Using the pressure gauge with 1/8” OD tubing, push tube into the fitting.

5. Turn the unit on in the [2] setting.

6. Make sure that the connection front the pressure gauge to the manifold is

leak-free. Part No. K1036-1 Pressure Test Kit (gauge and connector) and
KI340-1 Adaptor.

7. Put unit into test mode by pressing [1]

and [5] simultaneously for 10 seconds.

After 10 seconds, the alarm will sound and the LEDs above each flow setting
will illuminate. The unit will then start to automatically pulse in flow setting
[5].


Reference Section 6.1 Pressure Profiles